| Catalog Number |
1745G |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an eye procedure on (b)(6) 2018 and suture was used.The needle separated from the swage while passing through tissue with little tension.A like device of alternate product code was used to complete the procedure.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4).Investigation summary = > nine unopened samples of product code 1745g, lot llj546 were returned for evaluation.There were no defects found on the packets during the visual inspection.The samples were opened and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects were observed.A functional test was performed and pull force only the sample (# six) did not meet the requirements since, the result was below the minimum individual requirement, the eight other samples met the requirements.According to pull value results , the assignable cause is a needle pull out in the sample six.According to pull value results no performance pull off suture needle was found in eight samples.
|
| |
|