Catalog Number 1745G
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event or Problem Description
It was reported that the patient underwent an eye procedure on (b)(6) 2018 and suture was used.The needle separated from the swage while passing through tissue with little tension.A like device of alternate product code was used to complete the procedure.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Investigation summary
=
> nine unopened samples of product code 1745g, lot llj546 were returned for evaluation.There were no defects found on the packets during the visual inspection.The samples were opened and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects were observed.A functional test was performed and pull force only the sample (# six) did not meet the requirements since, the result was below the minimum individual requirement, the eight other samples met the requirements.According to pull value results , the assignable cause is a needle pull out in the sample six.According to pull value results no performance pull off suture needle was found in eight samples.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUT CHR UD 18IN 7-0 D/A CS175-8
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7476275
Report Number2210968-2018-72493
Device Sequence Number19372199
Product Code GAL
UDI-Device Identifier10705031002340
UDI-Public10705031002340
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number1745G
Device Lot NumberLLJ546
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Initial Date Received by Manufacturer 04/09/2018
Supplement Date Received by Manufacturer04/24/2018
05/24/2018
Initial Report FDA Received Date05/01/2018
Supplement Report FDA Received Date05/21/2018
06/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-