| Catalog Number |
1745G |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an eye procedure on (b)(6) 2018 and suture was used.The needle separated from the swage while passing through tissue with little tension.A like device of alternate product code was used to complete the procedure.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).An unopened representative sample was returned for analysis.During the visual inspection of sample, no defects were found on the package.The sample was opened and the swage and attachment area needles were as expected.The suture was dispensed without problems and examined along of the strand and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.Per the conditions of the sample received, no attachment defects were found and the tested sample met the finished goods requirements.
|
| |
|