| Model Number |
HC105-20 |
| Medical Device Problem Code |
Gradient Increase (1270)
|
| Health Effect - Clinical Codes |
Corneal Pannus (1447); Pulmonary Regurgitation (2023); Pulmonary Valve Stenosis (2024)
|
| Date of Event |
08/21/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: upon receipt at medtronic's quality laboratory, approximately 3.5 cm of the valve was returned with a transcatheter valve inside of the valve.Pannus fused the devices together.Due to the return condition of the product, the leaflets and commissures could not be assessed.Conclusion: medtronic is unable to determine a relationship between the product and the reported clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that four years and two months post implant of this bioprosthetic valve in a seven year old pediatric patient, a transcatheter pulmonary bioprosthetic valve was implanted valve-in-valve due to a gradient measurement of 90mmhg, regurgitation and stenosis.Prior to replacement, an echocardiogram (ecg) was performed that revealed the gradient measurement while the patient was exercising.No additional adverse patient effects were reported.
|
| |
|