| Model Number |
8888231316 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
04/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, when the patient returned to lying position after sitting up on bed, blood was noticed at the location of the catheter.It was found that the catheter was out and the suture wing remained attached to the skin of the patient.The intensive care unit (icu) and peripherally inserted central catheter (picc) nurse reported that the catheter came out of the patient.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted no abnormalities.Pmv performed functional testing which included submerging the catheter into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Both extensions were tested with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during treatment, when the patient assisted to supine position after trying to sit up on bed, it was noticed that blood was running out on the patient's chest.The catheter was placed in the right intra jugular site.It was found that the catheter was out and the suture wing remained attached to the skin of the patient.They held pressure on the access site to prevent further blood loss.It was reported that the patient did lose an amount of 100cc of blood.There was no blood transfusion.The patient was on day 2 of his treatment.The patient's vital signs remained stable and there were no reported signs of hypotension or air embolism.The treatment was not completed and the patient had to have another catheter placed.It was stated that the catheter was replaced on and the treatment was resumed.There was no reported patient injury.
|
| |
|