• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. BENCHMARK 6F 071 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number BMK6F95
Medical Device Problem Codes Leak/Splash (1354); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
During preparation for a medical procedure, the nurse removed a benchmark 6f 071 delivery catheter (benchmark) from the packaging and flushed it.The benchmark was then attached to a rotating hemostasis valve (rhv) and ready to use with no noted issue.However, when commencing continuous flush through the benchmark, the physician noticed that the benchmark was leaking around the hub.Therefore, the benchmark was not used in the procedure.The procedure was completed using a new benchmark.
 
Additional Manufacturer Narrative
Results: the benchmark was fractured underneath the strain relief approximately 2.5 cm from the hub.Conclusions: evaluation of the returned benchmark revealed that the catheter was fractured.This type of damage typically occurs if the device is forcefully retracted at extreme angles during removal from the packaging or general mishandling during preparation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK 6F 071 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7476947
Report Number3005168196-2018-00913
Device Sequence Number1769478
Product Code DQY
UDI-Device Identifier00814548014111
UDI-Public00814548014111
Combination Product (Y/N)Y
Initial Reporter CountryUK
PMA/510(K) Number
K142321
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/04/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/09/2020
Device Catalogue NumberBMK6F95
Device Lot NumberF78984
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/04/2018
Supplement Date Received by Manufacturer08/09/2018
Initial Report FDA Received Date05/01/2018
Supplement Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-