Product event summary: the sheath, 4fc12 with lot 50954, and the data files for the date of the reported event were returned and analyzed.The data files showed an unrelated system notice and low temperature and flow profiles.Visual inspection of the sheath showed that the shaft was kinked at 2.5 inches from the tip.It was noted that the kink was on the outer surface of the shaft.The kink was preventing a smooth steerability when the test balloon catheter was inserted.The shaft was out of plane when fully steered.In conclusion, the reported deflection and kink has been confirmed through testing.The sheath, 4fc12 with lot 50954, failed the performance test due to the shaft kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, the temperature was not as expected.The position of the balloon catheter was changed with resolve.Additionally, the balloon deflated after several seconds of treatment, and a system notice was received indicating that the safety system detected a compromised outer vacuum.It was also reported that the balloon shaft was broken.The balloon catheter was replaced and the case was completed with cryo.No patient complications have been reported as a result of this event.The sheath subsequently tested out of specification per the manufacturer's investigation.
|