• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® BARRICOR¿ LITHIUM HEPARIN PLASMAN BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® BARRICOR¿ LITHIUM HEPARIN PLASMAN BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365043
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field."multiple lot numbers: there were multiple lot numbers reported to be involved.The information for the additional lot number is as follows: medical device lot #: 6299942, medical device expiration date: 04/30/2018, returned to manufacturer on: 09/26/2017, device manufacture date: 10/25/2016".Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for underfilling with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.
 
Event or Problem Description
It was reported that the bd vacutainer® barricor¿ lithium heparin plasman blood collection tubes were underfilling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® BARRICOR¿ LITHIUM HEPARIN PLASMAN BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
MDR Report Key7477396
Report Number9617032-2018-02284
Device Sequence Number1592045
Product Code JKA
UDI-Device Identifier50382903650434
UDI-Public50382903650434
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Phlebotomist
Device Expiration Date11/30/2017
Device Catalogue Number365043
Device Lot Number6145857
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2017
Initial Report FDA Received Date05/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-