| Model Number |
D334DRM |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 4470 competitor lead, implanted: (b)(6) 2007.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was premature battery depletion and the device was at eri (elective replacement indicator).The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|