Model Number P20-0950
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pneumothorax (2012)
Date of Event 03/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 01-may-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.(b)(4).
 
Event or Problem Description
It was reported a pneumothorax was discovered after the nasogastric tube (ngt) placement.The x-ray revealed a lung pneumothorax.A chest tube was placed and a pigtail catheter was inserted to aid in resolving the pneumothorax.No additional information was provided.
 
Additional Manufacturer Narrative
One sampled device was received.All information reasonably known as of (b)(6) 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a halyard health product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The dhr noted the process has not changed for the manufacturing of the cortrak unit.The unit passed all required inspections and was working as expected prior release.1 cortrak unit that was returned with the original case shipper.The following components were returned with the monitor unit: sru - lot number 1212007, interconnect cable, battery pack, power cord.The monitor unit was inspected for damages and breakages, there were no breakages or damages found on the blue housing and orange button.Evaluation began by attaching the returned monitor unit, sru, interconnect cable to the lab's anatomical model, transmitting stylet.Functional testing of the unit indicated that the unit worked as intended.The smart receiver unit (sru) was able to communicate with the monitor unit, interconnect cable, transmitting stylet, anatomical model and performed self-test without any failures.The led on the sru was illuminating green all the time.The smart receiver unit was able to provide an accurate deployment of the transmitting stylet tip during the entire simulation placement, without any failures.After the setup of the cortrak system and the completion of the initialization and testing, the functionality of the unit performed as intended.The unit did not appear to be malfunctioning when the placement was taken, instead it appears to be lung placement that would have been apparent to a trained clinician.The evaluation tracing to the returned unit and the sru were able to provide an accurate deployment of the transmitting stylet tip during the entire placement, without any failures.For this reason, the root cause "user: incorrect use" was selected.The root cause was use error.All information reasonably known as of 27-jun-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
DH CORTRAK (EAS)
Manufacturer (Section D)
HALYARD HEALTH
5405 windward parkway
alpharetta GA 30004
MDR Report Key7477846
Report Number3006646024-2018-00016
Device Sequence Number11066912
Product Code KNT
UDI-Device Identifier10815149024698
UDI-Public10815149024698
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K113351
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other Health Care Professional
Device Model NumberP20-0950
Device Catalogue Number104720100
Device Lot Number1301034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/10/2018
Supplement Date Received by Manufacturer05/04/2018
06/01/2018
Initial Report FDA Received Date05/01/2018
Supplement Report FDA Received Date05/04/2018
06/28/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient Weight72
-
-