| Model Number |
DB-9218-15 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A report was received that the patient was experiencing impedance issues on one electrode of the m8 adapter.The physician replaced the adapter.
|
| |
|
Event or Problem Description
|
|
A report was received that the patient was experiencing impedance issues on one electrode of the m8 adapter.The physician replaced the adapter.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received that the patient did not have any symptoms as a result of the high impedance.The patient was doing well post-operatively.Analysis of db-9218-15 (s/n (b)(4)) revealed that the complaint has been confirmed.Visual (microscope) and x-ray inspection found a broken cable (contact #1) at the connector block.There are no exposed cables.The broken cable resulted in the reported complaint of high impedance.A review of the manufacturing documentation for the device revealed that no anomalies or deviations potentially related to the event occurred during manufacturing.
|
| |
|
Event or Problem Description
|
|
A report was received that the patient was experiencing impedance issues on one electrode of the m8 adapter.The physician replaced the adapter.
|
| |
|