| Model Number |
D334VRM |
| Medical Device Problem Codes |
Failure to Charge (1085); Electrical /Electronic Property Problem (1198); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patients implantable cardioverter defibrillator (icd) had triggered alerts for charge circuit timeout, charge circuit inactive, and long charge time prior to battery depletion.The device was extracted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated charge circuit inactive.Analysis of the device memory indicated charge circuit timeout.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed.Analysis of the device found high internal battery resistance.Device analysis confirmed that the device did not charge efficiently with the original device battery.The device charged nominally when the battery was replaced with an external supply.No hybrid anomalies were found.Battery analysis revealed that lithium-severing was seen in turn 7.The device was returned due to an end of service (eos), excessive charge time (ect) and charge time out (cto) alerts.This charge timeout resulted from increased battery resistance induced by observed lithium-severing.If information is provided in the future, a supplemental report will be issued.
|
| |
|