MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3341055
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.(b)(6).Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) patient underwent primary breast augmentation with a mentor siltex round moderate profile 125cc saline prosthesis.Rupture on the left breast prosthesis was diagnosed intra-operatively through a visual examination.As a result, the patient underwent unilateral removal and replacement with a memoryshape breast implant 225cc gel prosthesis on (b)(6) 2018.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed on 8/3/2018, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4),.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed on 6/18/2018, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The mentor failure analysis lab received the device for evaluation on 7/12/2018 and received updated information about the identity of the reported device: brand name: mentor memoryshape breast implant, procode: ftr, common device name: prosthesis, breast, noninflatable, internal, silicone gel-filled, catalog: 3341055, lot: 7515245, serial: (b)(4), unique identifier (udi): (b)(4), pma/ 510(k): p060028.A device history record review is in progress.Once completed, a supplemental report will be submitted.Additional information received indicates that the complaint device was a mentor leiden breast impant, but was mistakenly returned to mentor.The mentor leiden breast implants that are manufactured and distributed under the mentor brand but are not approved for import into the united states do not appear to meet the criteria for mdr reporting as a same or similar device to a device that is marketed in the united states.Specifically, the devices differ in manufacturing processes and device specifications.Events that involve these devices would not need to be reported as asrs/psrs or mdrs.As a result, no further analysis will take place.However, since this event has already been reported to fda, any additional information will continue to be reported as applicable.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation summary: the device was returned at mentor containing a clear gel.During initial evaluation a rent was observed on the anterior aspect, measuring approximately 9.7 cm.Microscopic examination of the edges of the rent gave no indications as to cause.No other anomalies were observed.Complaint was confirmed.Mentor performs 100% inspection and testing of all devices prior to release, the product evaluation team concluded that rents are sometime subsequent to the removal of the device from its protective packaging.The complaint was confirmed since a rupture was found on the device.Nonetheless, a microscopic examination of the edges of the rent was performed and it did not provide conclusive evidence of what could be the root cause.The device history record ( dhr) of lot number 7515245 reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Rupture is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7478477
Report Number1645337-2018-02558
Device Sequence Number1753819
Product Code FTR
UDI-Device Identifier00081317001713
UDI-Public00081317001713
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/12/2022
Device Catalogue Number3341055
Device Lot Number7515245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Initial Date Received by Manufacturer 04/09/2018
Supplement Date Received by Manufacturer04/09/2018
04/09/2018
04/09/2018
08/24/2018
Initial Report FDA Received Date05/01/2018
Supplement Report FDA Received Date06/25/2018
07/18/2018
08/07/2018
08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-