| Catalog Number |
TEXP140RH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab received the device for evaluation on (b)(6) 2018.The analysis has begun, but is not complete at this time.When the investigation and analysis have been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed on (b)(6) 2018, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old female patient underwent primary breast augmentation with a artoura breast tissue expander 600cc.Deflation on the left breast tissue expander was diagnosed post procedure.As a result, the patient underwent unilateral removal and replacement with a artoura breast tissue expander 600cc on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: it was reported that a (b)(6) year-old female patient underwent primary breast augmentation with a artoura breast tissue expander 600cc.Deflation on the left breast tissue expander was diagnosed post procedure.As a result, the patient underwent unilateral removal and replacement with a artoura breast tissue expander 600cc on (b)(6) 2018.The device was received at mentor with clear fluid.No foreign material was observed within the device or on the shell surface.The product evaluation team discovered a rent measuring approximately < 0.1cm on the anterior aspect.No other anomalies were discovered.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, the product evaluation team was precluded from further evaluating the device.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|