MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Imprecision (1307)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).On (b)(6) 2018, fse contacted the customer via phone to address the reported event.Inspection of the device revealed that the sample needle assembly was slightly bent and that one of the adapters over the trays was missing; resulting in the sample needle assembly being bent.The customer replaced the probe with another one he had on site.No further qc issues were noted.No further action was required by field service.A (b)(4) complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from (b)(6) 2017 through (b)(6) 2018.There were no similar complaints identified during the search period.The g8 service manual chapter 6 - assembly and disassembly (machinery) was reviewed.The 6.4.1 sampling needle replacement.The sampling needle replacement requires access to the inside of the analyzer.Be sure to wear protective clothing (goggles, gloves, mask, etc.) and take sufficient care to prevent infection during handling.Follow the procedure on page 6-5, "6.3.3 needle cover removal", and remove the needle cover.You will see the sampling needle unit back in the middle.Grasp the upper part of the sampling needle unit by hand and slowly pull the unit forward as much as possible.Since a small volume of reagent will spill during replacement, place a lab wipe under the sampling needle tip.By hand, loosen and remove the joint connected to the 3-way block.Remove the screws on the upper section of the sampling needle.Be careful not to drop the screws or the holding plate inside the machine during this operation.Remove the screws holding the stay, through which the tubing passes.Slowly lift up the sampling needle to remove.Insert the new sampling needle and secure the upper plate with the screws.When you do so, make sure that the needle tip hole is oriented forward.Pass the tubing through the stay, secure with the screw, and securely connect the joint to the 3-way block.Move the sampling needle unit back and forth and confirm that the tubing does not catch anything.If necessary, loosen the screws, turn the sampling needle, and change the stay direction to prevent the tubing from catching anything.Push the sampling needle unit back and attach the needle cover.Keep the following points in mind during replacement: be sure to wear protective clothing (goggles, gloves, mask, etc.) and take sufficient care to prevent infection during sampling needle replacement.Take care not to touch the end of the sampling needle.If the needle is bent immediately after replacement, check that the primary tubes match the sample rack or sample rack adapter.If the needle placement is clearly off center of the primary tube, it must be adjusted.Change the "y-samp" parameter.Refer to page 4-16, chapter 4.Dispose of the used sampling needle as infectious waste according to the procedures at your facility.The most probable cause of the reported event was due to a bent sampling needle assembly.
 
Event Description
On (b)(6) 2018, the customer reported that the g8 analyzer was not providing quality control (qc) results.The customer reported that qc had erratic peaks and no total area.Fresh qc was made the week prior and the day of the reported event at 1:110 dilution.The analyzer did provide results for patient samples.Technical support (ts) instructed the customer to make a dilution of the sample he ran that was normal; however, the issue did not resolve.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba- koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7478570
MDR Text Key107569235
Report Number8031673-2018-00396
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/01/2018
Distributor Facility Aware Date04/11/2018
Device Age5 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer05/01/2018
Initial Date Manufacturer Received 04/11/2018
Initial Date FDA Received05/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-