| Model Number |
G21393 |
| Medical Device Problem Code |
Kinked (1339)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The pig tail of the stent kinked easily when the customer straightened the pig tail using the straightener.The different device was used for the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining (b)(4).Importer site establishment registration number: (b)(4).Problem statement: the customer reported the following complaint issue: ¿the pig tail of the stent kinked easily when the customer straightened the pig tail using the straightener.The different device was used for the procedure.¿ additional information: the following additional information was provided: user's comment: i wonder if it's because the material was hard.Sales rep's comment: i think it was a defective device.Lab evaluation: 1 x zebd-7-4 device was returned to cirl for evaluation.A lab evaluation was held on 1st may 2018.During the lab evaluation, there was a kink noted at porthole 2.A 0.035 inch wire guide could not pass porthole 2.The kink was confirmed.There was indentation noted at porthole 5.There was no back wall damage noted.Root cause: a definitive root cause for the customer complaint could not be determined as the exact operational conditions of use could not be replicated in the laboratory setting.However, a possible cause of this complaint is that the stent may have been kinked while straightening the pigtails with the pigtail straightener.A precaution included in instructions for use states "care must be exercised when straightening the pigtail curls* in order to avoid kinking or breaking the stent." the complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Documents review: a review of the manufacturing records did not reveal any discrepancy related to the complaint issue.There is no evidence to suggest that this issue affects the entire lot # upon review of complaints this failure mode has not occurred previously with this lot number.Prd/fqc/pkg review: prior to distribution, all devices are subjected to 100% inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl as per fqc.As per fqc there is 100% inspection on all stents for kinks and a check that the appropriate size wire guide moves smoothly and freely when inserted into both ends of the stent.There is also a visual inspection of the product and packaging at packaging, packaging qc, and post sterile qc "complete a 100% visual inspection of the packaged unit (tyvek pouched) while held at a comfortable arm¿s length from the unaided eye at normal lighting conditions." ifu review: it may be noted that according to the instructions for use, the user is instructed to: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use¿.Summary: the complaint is confirmed as the failure was verified in the laboratory and the stent was found to be kinked.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
This follow up report is being submitted to updated the investigation results.Initial mdr submitted based on the device malfunction precedence: 'stent kinked/bent'.The pig tail of the stent kinked easily when the customer straightened the pig tail using the straightener.The different device was used for the procedure.
|
| |
|