| Model Number |
486619035 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
04/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while preparing the right l4, the pedicle finder tip broke off.The tip was retrieved using a clamp from a screw removal set.There was a 20 minute surgical delay.There are no reports of adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of device tip fracture was confirmed via visual inspection.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: the most likely cause of the reported event was determined to relate to : hard bone quality; previously damaged/deformed probe after long-term usage (about 9 years); excessive forces applied.
|
| |
|
Event or Problem Description
|
|
It was reported that while preparing the right l4, the pedicle finder tip broke off.The tip was retrieved using a clamp from a screw removal set.There was a 20 minute surgical delay.There are no reports of adverse consequences to the patient.
|
| |
|