• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT REV TIB 2 DEG BROACH 37MM; KNEE INSTRUMENT : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT REV TIB 2 DEG BROACH 37MM; KNEE INSTRUMENT : BROACHES Back to Search Results
Catalog Number 217863111
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Date of Event 04/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during the case, a few of the trials used did not effectively hold onto the stem trials.This did not create any major delays however twice the stem trial came loose and had to be retrieved from the patient's tibia.Following the case, i tested these trials and broaches and will be returning them to get new ones.The rubber gasket are getting worn out.There was a surgical delay of 5 minutes.
 
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).Investigation summary the device associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV TIB 2 DEG BROACH 37MM
Common Device Name
KNEE INSTRUMENT : BROACHES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7479142
Report Number1818910-2018-58702
Device Sequence Number1510415
Product Code HTQ
UDI-Device Identifier10603295096085
UDI-Public10603295096085
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number217863111
Device Lot NumberA0606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Initial Date Received by Manufacturer 04/09/2018
Supplement Date Received by Manufacturer06/13/2018
07/25/2018
Initial Report FDA Received Date05/02/2018
Supplement Report FDA Received Date06/14/2018
08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-