Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(4) 2018 in a follow-up call in order to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results the customer is concerned with results obtained of 144, 149, 147 and 170mg/dl.The expected fasting blood glucose test result range is 90-120mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 06/30/2019 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Customer does not have any more strips to trouble shoot and strips were stored incorrectly.
|
| |
|