Model Number FUSION |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
It was reported to manufacturer representative that the navigation system became unresponsive and shut down.This issue occurred outside of procedure.There was no patient present when this issue was identified.No additional information was provided.
|
|
Manufacturer Narrative
|
Correction: event date updated to proper value.The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.The behavior of the reported issue was found to be consistent with a known anomaly in the medtronic navigation software anomaly tracking database.This issue was documented in a medtronic navigation software anomaly tracking database.
|
|
Manufacturer Narrative
|
Unique device identification (udi) and manufacture date provided.If information is provided in the future, a supplemental report will be issued.
|
|