Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Visual Disturbances (2140); Halo (2227); No Code Available (3191)
Date of Event 04/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid, in a specimen cup.Visual inspection found no visible damage to the lens.(b)(4).Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm micl13.2, -12.50, implantable collamer lens was implanted into the patient's right eye (od) on (b)(6) 2018.On (b)(6) 2018 the lens was explanted due to observation of excessive vault, elevated intraocular pressure without pupil block and angle closure (date assessed: (b)(6) 2018), unreactive (fixed) pupil, glare/haloes, and blurred vision.Additional surgical intervention ( enlargement of pi; yag) and medical intervention (intraocular injection) were performed/prescribed.The reported stated that the cause of this event was unknown."fixed pupil" was reported for status of the eye at patient's last follow up.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information received: as of (b)(6) 2018 best corrected acuity 20/30, iop is elevated to 27um.Patient is being monitored for ocular hypertension.Corrections: patient did not have intraocular injections.(b)(4).
 
Additional Manufacturer Narrative
Device evaluation: lens was returned in liquid lens case/vial.Visual inspection found no visible damage to lens.Dimensional inspection found the lens within specifications.Corrected data: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7480272
Report Number2023826-2018-00657
Device Sequence Number16570223
Product Code MTA
UDI-Device Identifier00841542103343
UDI-Public00841542103343
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/31/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/04/2018
Supplement Date Received by Manufacturer05/08/2018
08/09/2018
Initial Report FDA Received Date05/02/2018
Supplement Report FDA Received Date06/05/2018
08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-