• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY EXACTA EXTERNAL DRAINAGE AND MONITORING SYSTEM; DEVICE, MONITORING, INTRACRANIAL PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY EXACTA EXTERNAL DRAINAGE AND MONITORING SYSTEM; DEVICE, MONITORING, INTRACRANIAL PRESSURE Back to Search Results
Model Number 46700
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Coma (2417); No Known Impact Or Consequence To Patient (2692)
Date of Event 04/27/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the device was leaking at one of the stopcocks.It was stated that it was unknown if there were any patient symptoms or complications associated with the event.The patient¿s medical history included trauma.The patient outcome was listed as alive ¿ with injury; they were currently in a coma.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional review indicated that the device was replaced due to the leaking.
 
Additional Manufacturer Narrative
The returned product was patent and met the requirements for leak testing.Therefore the conditions of this complaint could not be duplicated by laboratory personnel.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported the patient being unconscious was not related to the device.This was due to the patient¿s history of trauma.It was stated the leaking occurred at the main system stopcock.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTA EXTERNAL DRAINAGE AND MONITORING SYSTEM
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7480801
Report Number2021898-2018-00230
Device Sequence Number195988
Product Code GWM
UDI-Device Identifier00643169707665
UDI-Public00643169707665
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K983799
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2022
Device Model Number46700
Device Catalogue Number46700
Device Lot Number214690877
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2018
Supplement Date Received by Manufacturer05/02/2018
05/22/2018
05/16/2018
Initial Report FDA Received Date05/02/2018
Supplement Report FDA Received Date05/02/2018
05/24/2018
05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age69 YR
Patient Weight65
-
-