| Model Number |
705015 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that during an axsos proximal humerus case, the screwdriver bit which is attached to the torque limiting adaptor was used to insert the screw by hand.Prior to achieving 2.5nm torque, the tip of the screwdriver broke off.It was reported that the broken tip remained inside the screw-head.It was reported that the screw was left in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that screwdriver bit t15, ao fitting was alleged of 'breakage during surgery' could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by to much applied mechanical force during screw insertion.The device inspection revealed the following: the returned screwdriver bit is indeed completely broken off.The close up of the fracture surface show some cracks (see attached image documentation) whereas the high applied forces led to a strain hardening / embrittlement of the material which finally resulted in the breakage.The root cause of the failure was repeated powerful dynamically overload during use.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an axsos proximal humerus case, the screwdriver bit which is attached to the torque limiting adaptor was used to insert the screw by hand.Prior to achieving 2.5nm torque, the tip of the screwdriver broke off.It was reported that the broken tip remained inside the screw-head.It was reported that the screw was left in the patient.
|
| |
|