| Model Number |
ADVIA 1800 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A siemens customer service engineer (cse) was dispatched to the customer site.The cse observed that the incorrect calibration factor (cf) was being used on the instrument for the ca_2 assay.The calcium (ca) cf factor was being used instead of ca_2 factor.The cse changed the calibration factor on the instrument and noted that the quality control values dropped closer to the mean.The cause of the falsely high ca_2 values was due to the incorrect factor being used on the instrument.The device is performing within manufacturing specifications.No further evaluation of the device is required.
|
| |
|
Event or Problem Description
|
|
Discordant, falsely elevated (18-20% higher) calcium arsenazo (ca_2) results were obtained on patient samples on an advia 1800 instrument.It is unknown which reagent lot was used.The customer has declined to provide any patient sample data to siemens.It is unknown if the samples were repeated and if the initial and/or repeat results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated calcium arsenazo results.
|
| |
|