• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that after the lens was inserted into the patient's right eye, two large scratches were observed across the optic.The lens was then removed and replaced with another lens of the same model and diopter without incident.There was no patient injury and no surgical or medical intervention required.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 5/2/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eyes shows the product is cut in half, most probably to make remove and replacement possible.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7482157
Report Number3011852734-2018-00070
Device Sequence Number1630463
Product Code MFK
UDI-Device Identifier05050474560666
UDI-Public(01)05050474560666(17)210507
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/07/2021
Device Model NumberZLB00
Device Catalogue NumberZLB00U0195
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Initial Date Received by Manufacturer 04/05/2018
Supplement Date Received by Manufacturer05/31/2018
Initial Report FDA Received Date05/02/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL 1MTEC30; CARTRIDGE MODEL 1MTEC30
Patient Age71 YR
-
-