On (b)(4) 2018, iba was made aware of a potential event in the clinical workflow of one customer in (b)(6), resulting in the correction vector not being applied before patient treatment.
This event occurred after an upgrade of the hand pendant from wired to wireless at end of (b)(6) 2018.
A number of patients has potentially received their treatment at a wrong position.
The client and iba are currently investigating the possible impact on the treatment of each individual patient for the period in question.
Since (b)(6) 2018, the client has adapted the clinical workflow to ensure that the correction vector is applied.
|