Catalog Number UNKNOWN |
Device Problems
Difficult to Remove (1528); Unintended Movement (3026); Insufficient Information (3190)
|
Patient Problems
Dyspnea (1816); Pain (1994); Palpitations (2467); No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook celect filter.Name and address for importer site: (b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2014".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).F10: patient code: dyspnea (1816), not listed in ifu; pain (1994), not listed in ifu; palpitations (2467), not listed in ifu.Device code: unintended movement (3026) [device tilt], listed in ifu; difficult to remove (1528), not listed in ifu.G1) name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
This additional information received on 17apr2018 as follows: pt allegedly received an implant on (b)(6) 2014 due to pulmonary embolism.Pt alleges tilt, device is unable to be retrieved.Pt further alleges shortness of breath, heart palpitations, right clavicle/shoulder pain and pressure, parasites and weakness.Attempted retrieval on (b)(6) 2015 was unsuccessful.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).G1) name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tilt, unable to retrieve, sob, heart palpitations, shoulder pain/pressure, parasites, weakness".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Unknown if the reported shoulder pain/pressure, shortness of breath are directly related to the filter.Unknown if the reported heart palpitations, parasites, weakness are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|