• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. AGILIS¿ INTRODUCER, UNKNOWN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. AGILIS¿ INTRODUCER, UNKNOWN Back to Search Results
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
A leak was noted at the hemostasis valve during functional testing.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.In addition, the sheath deflected in both directions when actuating the steering mechanism, and deflected in the correct shape with no resistance noted.The cause of the torn seals and subsequent leak is consistent with damage during use.The ifu states: do not remove dilator or catheter rapidly.Damage to the valve may occur, potentially compromising hemostasis.
 
Event Description
This report is to advise of an event observed during analysis confirming a hemostasis valve leak of the introducer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS¿ INTRODUCER, UNKNOWN
Type of Device
INTRODUCER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7482536
MDR Text Key107197767
Report Number2182269-2018-00065
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-