| Model Number |
0010-6110 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product evaluation: failure analysis for fiber 0010-2090-721h-(b)(4): the fiber does not exhibit signs of usage; the fiber is broken within the white tubing at 8.25 inches from control knob, between the connector and control knob; the fiber was tested with hene laser fixture, aim beam is present at location of the break; the white tubing does not exhibit signs of melting at the break.Review of lot 721h dhr did not indicate any failures upon the final inspection.Probable root cause: based on the device analysis, the probable root cause of the failure could not be established based on the results of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgical procedure, the fiber did not work.The fiber was exchanged and the procedure was completed utilizing a second fiber.Patient outcome: no injury to the patient was reported.
|
| |
|