| Model Number |
73711 |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/04/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No evaluation possible at this time.The instrument has not been returned for evaluation.The illico awl is a cannulated instrument ground at 15° to form a sharp trocar tip intended to penetrate cortical bone during pedicle preparation.The section retained by the patient is manufactured from custom (b)(4) stainless steel which is processed and certified to astm (b)(4) specifications.
|
| |
|
Event or Problem Description
|
|
The distal tip of an illico awl broke off during the mis case.The detached section remains captured within the patients l5 pedicle.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the returned device confirmed the distal tip, approximately.591 in length has been completely separated/detached from the instrument.Under magnification, the fracture patterns reveal lateral forces where applied to the instrument which caused the ø.20" tip to encounter excessive fatigue/stress.
|
| |
|