| Model Number |
48770410 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It is reported that fixation pin for the plate broke at the distal end during screwing in.The distal end stayed into the patient.The round head part of the screw broke apart.New screw was implanted.Surgical delay of 30min is reported.No adverse consequences is reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One fixation pin was reported to have broken in the threaded area while attempting to thread the pin in the patient's bone.The device was returned for revaluation, manufacturing history was reviewed and no anomalies were found.Material analysis was performed and the submitted fixation pin was found to have fractured in a mode of torsional ductile overload.The elemental constitution was as specified on the drawing.No material or manufacturing defects were found.The sales representative reported that the patient had hard bone quality which is the likely cause of the tip fracture.
|
| |
|
Event or Problem Description
|
|
It is reported that fixation pin for the plate broke at the distal end during screwing in.The distal end stayed into the patient.The round head part of the screw broke apart.New screw was implanted.Surgical delay of 30 min is reported.No adverse consequences is reported.
|
| |
|