| Model Number |
8886302361 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a patent foramen ovale/atrial septal defect endoscopic procedure, while making the purse string suture, the needle detached from the thread.A new suture was used in order to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned noted one suture and one needle.The needle was broken at the swaged end.The swaged end of the needle was attached to the suture.It was observed, under magnification, that instrument marks were present on the swaged end of the needle.The retainer was inspected with no abnormalities.As no sealed products were returned, functional testing was precluded.Because the broken swaged end of the needle is attached to a suture, the reported condition was not confirmed.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.The observed secondary condition of needle broken suggested that the needle was grasped improperly at the swaged end of the needle during application.Firmly grasping the needle at the swaged end can deform the crimp, can cause the needle to disengage from the suture, and can even cause the needle to break at the swaged end.The recommended grasping area of the needle is at the needle body, where the wire is of a full diameter.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|