| Model Number |
M004PM4790K20 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Bent (1059)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: analysis of returned product revealed the unit passed electrical and rf ablation testing, consequently, not confirming the reported complaint of "noisy electrogram"; however, the device had a kink at the distal section-approximately the distal section while in the neutral position-ring 1 seals were compromised; there was dried body fluid on the edge of the electrode.Bent center support and a steering wire partially detached from the center support were observed under x-ray between r1 and r2 at the distal section.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
This event is reportable based on device analysis completed 4/6/18.It was reported that upon ablation in the right atrium, noise appeared on the 1-2 bipolar display.It was necessary to replace the catheter to complete the procedure.No patient complications occurred.It was also noted that the tip of the catheter was bent out of the box.Device analysis revealed a compromised seal on ring 1.
|
| |
|