Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Device history record review revealed nothing relevant to this event.The evaluation revealed that the returned bio-composite screw is degraded.The condition reported may be related to a patient reaction to the materials implanted.Absorption of biodegradable implants begins at implantation as a result of normal hydrolysis of the implant.The rate at which the implant degrades and is reabsorbed can be affected by numerous factors including, but not limited to implantation site vascularity, stress on the implants, surgical techniques employed during implantation, patient factors (such as age, immune system status, bone quality, body chemistry and sensitivity to the implant.The potential causes of this event are being communicated to the event reporter.If the device is returned and additional information is obtained, a follow-up report will be submitted.
|