| Model Number |
8884741220 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer stated that the tube was broken when opened by the radiologist before putting into patient, prior to priming.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record file was reviewed indicating that product was released meeting all quality standard requirements.A sample was received.After a visual inspection was performed, a tube was found to be broken.The reported issue was confirmed.A root cause could not be determined.A formal corrective and preventative action has been opened to address this issue.The investigation will be used for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
|
| |
|