On (b)(6) 2016, a crown cna27 aortic heart valve prosthesis was implanted.No antibiotics were used prior to the valve implant.On (b)(6) 2018, the valve was explanted due to prosthetic endocarditis and abscess.The valve was replaced with another crown cna27.The valve was functioning properly prior to the explant.Only slight sclerosis of the valve was identified.No signs of fever were identified pre-operatively.A blood culture test was performed and resulted positive for staphylococcus hominis.No previous patient endocarditis recorded.During explant a huge vegetation was identified as well as huge annular abscess of the valve.The patient outcome is recorded as a good outcome.
|
|
Correction: please note.In the original submission the event description (event) stated the date of implant was on (b)(6) 2018.This has been corrected in the updated event description to on (b)(6) 2016, along with the inclusion of further information received may 7, 2018.Manufacturer narrative: if staphylococcus was present on the tissue valve before sterilization, it would be killed very easily by the manufacturers liquid chemical sterilant.The sterilant contains a mixture of glutaraldehyde, formaldehyde and alcohol, all of which are highly effective against staphylococci.The manufacturer has validated this liquid chemical sterilization process with spore forming bacillus atrophaeus, which is the most resistant microorganism known for aldehydes.In addition, the sterile packaging solution used for shipment of crown valves is a (b)(4) formaldehyde solution; the valves remain in this extremely antimicrobial solution until they are opened during surgery.Therefore, it is not conceivable that staphylococci could survive on a valve exposed to the manufacturers sterilant, or the packaging solution.In addition the manufacturing and material records for the crown prt aortic pericardial heart valve, model#: cna27, s/n#: (b)(4) , as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova canada corp.The results confirmed that this valve satisfied all sterilization, material, visual, and performance standards required for a crown prt aortic pericardial heart valve at the time of manufacture and release.Due to the above reasoning, the large amount of time that had elapsed between implantation and the reported endocarditis, and the proper functionality of the valve during earlier follow-ups, the root cause is determined to be not valve related.The infection therefore could not be resultant of any factor intrinsic to the device in question and the device failure is related to the patient condition.Fields changed: date of report, event, date rec¿d by mfr, if follow-up, what type.
|