| Model Number |
ZXR00 |
| Medical Device Problem Code |
Dent in Material (2526)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was dimple on the lens.There was no patient contact reported.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 5/1/2018.Device returned to manufacturer ¿ yes device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.The product was received in a lens holder.The lens was inspected by a qualified final inspection operator using 12x magnification.There were no anomalies.The customer''s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|