MPRI ATTAIN OTW; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE Back to Search Results
Model Number 419378
Medical Device Problem Codes Failure to Capture (1081); Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: dtba1d1 icd, implanted: (b)(6) 2014; 5076-52 lead, implanted: (b)(6) 2002.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient's "wire broke" and "they are mad that a second one has done this".It was subsequently reported that the left ventricular lead was unable to capture at high outputs, had high and undefined impedance and a possible fracture.The lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the proximal portion of the lead was returned, analyzed, and no anomalies were found.The outer insulation of the lead was observed to have blood ingression.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN OTW
Common Device Name
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7484942
Report Number2649622-2018-07597
Device Sequence Number1630434
Product Code LWP
UDI-Device Identifier00885074200729
UDI-Public00885074200729
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P010015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/29/2008
Device Model Number419378
Device Catalogue Number419378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Initial Date Received by Manufacturer 04/05/2018
Supplement Date Received by Manufacturer07/27/2018
Initial Report FDA Received Date05/03/2018
Supplement Report FDA Received Date08/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694758 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient Weight63
-
-