• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC. THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D132701, PMA # P030031/S053

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC. THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; SIMILAR DEVICE D132701, PMA # P030031/S053 Back to Search Results
Lot Number UNKNOWN
Medical Device Problem Codes High Sensing Threshold (2574); Impedance Problem (2950)
Health Effect - Clinical Codes Cardiac Arrest (1762); Cardiac Tamponade (2226); Not Applicable (3189)
Date of Event 04/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No device was received for analysis at the time of submission of the initial 3500a.Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.Device history record (dhr) review cannot be conducted because no lot number was provided by the customer.As such the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.(b)(4).
 
Event or Problem Description
It was reported that a (b)(6) male patient underwent an ablation procedure for ventricular tachycardia with a thermocool® smart touch® sf bi-directional navigation catheter and suffered a cardiac arrest (requiring defibrillation) and a cardiac tamponade (requiring pericardiocentesis and surgical intervention).Pre-procedure, the patient was admitted to the facility post-myocardial infarction and ventricular tachycardia.Post-admission, the patient was intubated.At an unspecified point, the patient went into ventricular tachycardia and ventricular fibrillation and received greater than 30 defibrillator shocks (icd and external).During the procedure, there were 2 acute increases in impedance and high force readings on the force graph (but not on the dashboard).After approximately 60 minutes of ablation, the patient became hypotensive and a pericardial effusion was detected.Pericardiocentesis yielded an unspecified volume of fluid.Patient required an open chest surgical intervention to repair a tear in the cardiac tissue.There is no information regarding extended hospitalization or patient outcome.Physician indicated that the adverse event was secondary to catheter manipulation and the number of shocks delivered.It was noted that the smarttouch catheter was not at fault.Since this adverse event required medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure, it is to be considered serious and mdr reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Common Device Name
SIMILAR DEVICE D132701, PMA # P030031/S053
Manufacturer (Section D)
BIOSENSE WEBSTER INC.
33 technology drive
irvine CA 92618
MDR Report Key7485271
Report Number2029046-2018-01513
Device Sequence Number1456510
Product Code LPB
UDI-Device Identifier10846835010169
UDI-Public10846835010169
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P030031/S053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/09/2018
Initial Report FDA Received Date05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age65 YR
-
-