Model Number 980
Medical Device Problem Code Device Inoperable (1663)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer (fse) inspected the device and could not duplicate the reported condition.The ventilator has passed performance verification testing (pvt), extended self test (est), and short self test (sst) and is ready for use.The ventilator has been returned to the customer.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient, a 980 ventilator shut down.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key7485637
Report Number8020893-2018-00207
Device Sequence Number13817152
Product Code CBK
UDI-Device Identifier10884521171565
UDI-Public10884521171565
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K162738
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980X1ENASAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2018
Initial Report FDA Received Date05/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-