Model Number 48-30-00
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova (b)(4) received a report that on a s5 system the display locks up intermittently during priming.There was no patient involvement.
 
Additional Manufacturer Narrative
Previously, complaints related to the firmware were considered to be reportable malfunctions.However, as the display and control module is part of the system panel, which is the interface between the user and the perfusion system, the module only displays sensor data, time measurements and status messages.It is partially used for the alarm management, configuring and assigning monitoring functions and providing information to the ups module.The module is not used for configuring and controlling the pumps themselves.The reported malfunction would not have any affect on device performance or the ability to maintain blood flow during cardiac surgery.Consequently, livanova (b)(4) was unable to identify a hazardous situation potentially leading to a serious injury as a result of this type of issue, and it is now considered to be non-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7485966
Report Number9611109-2018-00946
Device Sequence Number2052792
Product Code DTQ
UDI-Device Identifier04033817900887
UDI-Public010403381790088711170914
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K071318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48-30-00
Initial Date Received by Manufacturer 04/05/2018
Supplement Date Received by Manufacturer05/23/2018
Initial Report FDA Received Date05/04/2018
Supplement Report FDA Received Date06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-