• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE; FCG, BIOPSY NEEDLE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE; FCG, BIOPSY NEEDLE KIT Back to Search Results
Catalog Number ECHO-3-22
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Date 04/07/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use include the following information to ensure proper use of the device: "needle must be retracted into sheath and thumbscrew on safety ring must be locked to hold needle hold needle in place prior to introduction, advancement or withdrawal of device.Failure to retract needle may result in damage to endoscope." "introduce ultrasound needle into accessory channel and advance device in small increments until luer lock fitting at base of sliding sheath adjuster meets fitting on accessory channel.Attach device to accessory channel port by rotating device handle fittings are connected." "with ultrasound endoscope and device straight, adjust needle to desired length by loosening thumbscrew on safety ring, and advancing it until desired reference mark for needle advancement appears in the window of safety ring.Tighten thumbscrew to lock safety ring in place.Note: number in safety ring window indicates extension of needle in centimeters.Caution: during needle adjustment or extension, ensure device has been attached to accessory channel.Failure to attach device prior to needle adjustment or extension may result in damage to endoscope." bends and kinks in the sheath can occur if the device experiences excessive pressure.Prior to distribution, all echotip ultra endoscopic ultrasound needles are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
Prior to an endoscopic ultrasonography (eus), the physician selected a cook echotip ultra endoscopic ultrasound needle.The echo tip needle was bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE
Type of Device
FCG, BIOPSY NEEDLE KIT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key7486276
MDR Text Key107344536
Report Number1037905-2018-00179
Device Sequence Number1
Product Code FCG
UDI-Device Identifier00827002315211
UDI-Public(01)00827002315211(17)201122(10)W3997286
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K171596
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberECHO-3-22
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/07/2018
Device Age5 MO
Event Location Hospital
Initial Date Manufacturer Received 04/10/2018
Initial Date FDA Received05/04/2018
Date Device Manufactured11/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-