| Model Number |
CIE3-12-24-00 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Off-Label Use (1494); Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Pain Relief (2388)
|
| Date of Event |
04/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Our investigation of this event with the information to date suggests that the event (migration and revision) was due to off-label use of the varilift-lx device in a patient with grade 2 or 3 spondylolisthesis.The indications for use and all labeling for the varilift-lx informs users that this device is for use in patients with up to a grade 1 spondylolisthesis.There is no indication of device failure or malfunction.The varilift-lx device was not removed from the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a removal of a varilift-lx device was attempted due a reported migration in a patient with grade 2+ spondylolisthesis.
|
| |
|