• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PERSONAL CATHETER¿ INTERMITTENT CATHETER; HYDROPHILLIC PERSONAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PERSONAL CATHETER¿ INTERMITTENT CATHETER; HYDROPHILLIC PERSONAL CATHETER Back to Search Results
Model Number 63612
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the catheter lacked lubricity.The patient stated that the catheter may have not been well coated due to the catheter being coated for less than 25-30 seconds.
 
Additional Manufacturer Narrative
The reported event was unconfirmed.Visual evaluation noted no obvious defects.5 catheters were tested and the lubricious coating remained after 10 wipes.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿personal catheter intermittent catheter hydrophilic coated silicone instructions for use intended use: the catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/ precautions and adverse reactions.The catheter becomes slippery when wetted with water, eliminating the need for a separate lubricant.For your added convenience, this catheter is packaged with its own sterile water.Simply release the water from its foil packet and then tip the un-opened catheter package end-to-end.The catheter acts like a magnet to attract the water and activate its slippery coating.Follow these steps for best results.1.Release the water prior to opening the sealed catheter pouch: a.Apply pressure to the foil packet to release the water.B.Ensure all water is released from the foil packet.2.Wet the catheter a.Hold package with printed side up.B.Tip package end-to-end three to six times to wet catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.3.Use the catheter a.Peel back package to expose funnel end of catheter.B.If desired, use the self-adhesive tape on the package to temporarily attach the pouch to any dry vertical surface.C.Remove catheter and use according to physician¿s instructions.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Precautions: urethral catheter for urological use only.Discard after use, do not resterilize.Made of silicone elastomer.Manufacturer catalog number lot number units caution: federal (usa) law restricts this device to sale by or on the order of a physician.Single use.Not made with natural rubber latex.Not made with pvc.Sterilized using irradiation.Do not use if package is damaged.Use by manufacturer: c.R.Bard, inc.(b)(4) caution: consult accompanying documents.".
 
Event or Problem Description
It was reported that the catheter lacked lubricity.The patient stated that the catheter may have not been well coated due to the catheter being coated for less than 25-30 seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONAL CATHETER¿ INTERMITTENT CATHETER
Common Device Name
HYDROPHILLIC PERSONAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7487621
Report Number1018233-2018-01516
Device Sequence Number19160983
Product Code KOD
UDI-Device Identifier00801741069284
UDI-Public(01)00801741069284
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2022
Device Model Number63612
Device Catalogue Number63612
Device Lot NumberJUBR0258
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2018
Initial Date Received by Manufacturer 04/13/2018
Supplement Date Received by Manufacturer07/23/2018
Initial Report FDA Received Date05/04/2018
Supplement Report FDA Received Date08/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-