• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012452-08
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspections were performed on the returned device.The inner member/outer member was noted to be stretched at the proximal seal, fluid did not pass through noted stretched outer member, thus the reported inflation issue was confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported inflation issue.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of this device.
 
Event or Problem Description
It was reported the procedure was to treat a lesion in the moderately calcified, proximal left anterior descending coronary artery.The 3.75x08 mm nc trek rx balloon dilatation catheter (bdc) was advanced without issue in the patient anatomy.However, the balloon completely failed to inflate.The bdc was removed from the patient anatomy without issue and a same size nc trek was used to complete the procedure.There were no adverse patient effects.There was no clinically significant delay in the procedure.Returned device analysis identified the proximal seal was stretched.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7487644
Report Number2024168-2018-03318
Device Sequence Number10811618
Product Code LOX
UDI-Device Identifier08717648152016
UDI-Public08717648152016
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Catalogue Number1012452-08
Device Lot Number70906G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2018
Initial Report FDA Received Date05/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-