• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED CUTTER 20G/8000CPM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DUTCH OPTHALMIC RESEARCH CENTER BV DISPOSABLE HIGH SPEED CUTTER 20G/8000CPM Back to Search Results
Catalog Number 8267.VIT20
Medical Device Problem Code Device Operational Issue (2914)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/02/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation showed that the o- ring , the stainless steel ring and the plastic ring are not straight against the body, therefore the seal between the inner knife and the body can not be guaranteed to be closed.Vitrectomy produces air bubbles that can lead to injury by microbes and the risk is injury by microbial, viral or prion infection (patient harm).This reportable event was identified during a voluntary retrospective review of all complaints since 2015 by the manufacturer (dutch ophthalmic (b)(4)) details of this activity were discussed with cdrh office of compliance (b)(4) during a tele-conference on 05/31/2017.All available information has been disclosed.This complaint has now been closed.
 
Event or Problem Description
This is to report an event that occurred in the (b)(6).The victrectoom spits on the occurrence and does not suck well.Vitrectoom spit bubbles.Vitrectoom always gives air to the system.There is no further information and all information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE HIGH SPEED CUTTER 20G/8000CPM
Common Device Name
DISPOSABLE HIGH SPEED CUTTER 20G/8000CPM
Manufacturer (Section D)
DUTCH OPTHALMIC RESEARCH CENTER BV
scheijdelveweg 2
zuidland, 3214V N
NL  3214VN
MDR Report Key7488475
Report Number1222074-2018-00098
Device Sequence Number13757168
Product Code HQE
UDI-Device Identifier08717872019093
UDI-Public08717872019093
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K142877
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue Number8267.VIT20
Device Lot Number2000366703
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2015
Initial Report FDA Received Date05/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-