| Catalog Number |
72203994 |
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); High Blood Pressure/ Hypertension (1908)
|
| Date of Event |
04/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the dr.Had connected machine to the patient and started surgery by creating pneumoperitoneum with the insufflator.After some time they noticed that the patient's pressure had increased very high and also noticed bleeding from peritoneal layer.When they looked at the insufflator pressure, it was showing 12 mmhg actual pressure as same as set pressure but gas was flowing continuously.Immediately they stopped flow and with great effort they stopped bleeding.They finished surgery with standby machine.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, a 72203994 - smith & nephew 500 insufflator serial number (b)(4) , was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed and the product number and serial number were verified.No deficiencies were observed.No co2 input cap or accessory or ancillary products were returned with the device.A functional evaluation did not reveal any problems.The unit powered up and no error messages occurred.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.Please refer to the smith & nephew 500 insufflator operator's manual for troubleshooting suggestions and probable causes of problems encountered.The device has been sent to the original equipment manufacturer (oem) for further evaluation.
|
| |
|
Search Alerts/Recalls
|