| Catalog Number |
214140 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep that the jaw on the customer's expressew broke off of the device while inspecting it on the back table during a shoulder scope.The sales rep also stated that two expressew iii needles became stuck in the gun due to the clear tab on the needle coming off.The case was completed with another expressew iii gun and a different lot of expressew iii needles.There were no patient consequences but a 2 minute delay was reported to switch out the devices.No parts of the expressew iii gun fell into the patient.The expressew iii needles were discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A device history record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|