| Catalog Number |
71210003 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Box of pins was opened to reveal one pin had pierced plastic protective end and through plastic package as well.Scrub nurse was pricked by exposed sharp end.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated complaint devices and packaging were returned and evaluated.A visual examination of the returned trocar pins and packaging confirmed that stated complaint.One of the pins punctured the protect covering.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A clinical analysis indicated that several attempts have been made to obtain clinical/medical documents to no avail.Without supporting clinical/medical documents a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.A review of complaint history on the listed part revealed no prior complaints for the listed batch.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|