• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAL
Device Problems Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin squirted out during manual prime process.The customer¿s blood glucose was 88 mg/dl.Troubleshooting was performed.The customer also stated that they were receiving an a33 alarm.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Insulin pump passed the displacement test but compromised forced sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to compromised forced sensor system alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7489794
MDR Text Key107453046
Report Number3004209178-2018-69434
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503694
UDI-Public(01)00643169503694
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 09/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Device Lot NumberA4751NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/12/2018
Initial Date FDA Received05/07/2018
Supplement Dates Manufacturer Received08/14/2018
Supplement Dates FDA Received09/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
Patient Weight76
-
-