Model Number MMT-751NAL |
Device Problems
Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin squirted out during manual prime process.The customer¿s blood glucose was 88 mg/dl.Troubleshooting was performed.The customer also stated that they were receiving an a33 alarm.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Insulin pump passed the displacement test but compromised forced sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to compromised forced sensor system alarm.
|
|