MAUDE Adverse Event Report: COOK INC TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Model Number G23905
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 04/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a leak was found on a turbo-ject® double lumen over-the-wire power-injectable picc at the proximal tip after connection to the pump.A new picc line was necessary; therefore, implantation of the new line was completed which resulted in an extension of their hospital stay.
 
Additional Manufacturer Narrative
Additional information: investigation - evaluation a review of the complaint history, device history record, documentation, instructions for use (ifu), specifications, quality control, and visual inspection of the returned device was conducted during the investigation.One device was returned for investigation.A visual examination of the returned catheter noted the catheter had partially separated from the purple hub.Quality control specification confirm correct extension tube with winged flla and that they are securely attached.Inspection method: level i, measure tube outside diameter with calipers or micrometer.Visually examine.Manually tug.The complaint is confirmed.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time, however appropriate measures have been initiated to address this failure mode.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event for continuing consideration.
 
Event or Problem Description
It was reported a nurse noticed there was a leak of the turbo-ject® double lumen over-the-wire power-injectable picc located right below the luer-lock part of one of the two lumens.The peripherally inserted central catheter (picc) was removed and replaced.No other adverse consequences to the patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT® DOUBLE LUMEN OVER-THE-WIRE POWER-INJECTABLE PICC
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7489797
Report Number1820334-2018-01328
Device Sequence Number18859247
Product Code LJS
UDI-Device Identifier00827002239050
UDI-Public(01)00827002239050(17)191121(10)8389811
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K132885
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberG23905
Device Catalogue NumberUPICDS-5.0-CT-OTW-ST-1111
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/17/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date05/07/2018
Supplement Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-